Wheatstone bridge simulator

A Wheatstone bridge measures resistance very precisely by balancing two voltage dividers against each other. Adjust R3 until the meter reads zero, then work out the unknown.

Intermediate

meter

Challenges

  1. Balance the bridge with the fixed resistor as Rx.
  2. Measure the mystery resistor by balancing the bridge.
  3. Set the PT100 to 100 °C and balance the bridge.
  4. Balance the PT100 at 25 °C with no more than 1 mA through it.

Controls

Readings

What's going on

The bridge is two voltage dividers side by side. The left one, R1 over R2, sets one voltage. The right one, R3 over Rx, sets another. The meter in the middle shows the difference. When the two ratios are equal, both middle points sit at the same voltage, the meter reads zero, and the bridge is balanced:

R1 ÷ R2 = R3 ÷ Rx    so    Rx = R3 × R2 ÷ R1

The supply voltage doesn't appear in that answer, and at balance no current flows through the meter, so neither the supply nor the meter's own resistance affects the result. That's why bridges can be so accurate. Samuel Hunter Christie described the circuit in 1833, and Charles Wheatstone made it popular in 1843.

Where bridges are used

Strain gauges, load cells in weighing scales, pressure sensors and resistance thermometers are all read with bridges. Often the bridge isn't balanced at all: it's left slightly out of balance and the small output voltage is amplified, because that voltage changes with the quantity being measured.

The PT100 sensor

A PT100 is a platinum resistance thermometer that measures exactly 100 Ω at 0 °C. Its resistance rises with temperature following the IEC 60751 equation R = 100 × (1 + A·T + B·T²) from 0 °C up, where A = 3.9083 × 10⁻³ and B = −5.775 × 10⁻⁷. At 100 °C it measures 138.51 Ω. The current through the sensor warms it slightly, so bridges are usually designed to keep it to around 1 mA or less.

How this simulation works

The resistors and supply are ideal and the meter draws no current, so the maths is exact. The bridge counts as balanced when the meter reads within 0.1% of the supply voltage. The PT100 follows the IEC 60751 equation exactly. The model ignores self-heating and the resistance of the sensor's leads, which real installations handle with three-wire or four-wire connections. The mystery resistor is a random standard E24 value between 100 Ω and 9.1 kΩ, chosen again each time you load the page.